Specifications

Measurement time per point CW mode, at 500 kHz IF bandwidth < 4.5 μs
Measurement time
(including any data transfer time)
201 measurement points,
measurement bandwidth 500 kHz, display off,
ALC off, at 5 GHz, narrow span
< 4.5 ms
Data transfer time   data transfer simultaneous with measurement
Dynamic range at 10 Hz IF bandwidth > 120 dB
Inherent noise   < –110 dBm
Power sweep range1)   –40 dBm to +13 dBm
IF bandwidths   1 Hz to 500 kHz
Number of measurement points per trace   1 to 60 001
Platform    
Number of test ports   2 or 4
Number of measurement and reference receivers   one measurement and one reference receiver per
test port
Number of integrated generators   one generator per test port pair
Calibration techniques   TOSM, UOSM, TRL/LRL, TOM, TRM, TNA,
automatic calibration unit
Operating system and internal PC   Windows XP
Operation   hardkeys/softkeys, keyboard and mouse, online
help system, measurement and calibration wizard
Number of traces, diagrams, independent channels, setups that can be simultaneously loaded into RAM   > 1002), traces can be assigned to diagrams as
required

1) Power level that can be electronically swept.

2) Number limited only by internal memory.